Introduction
We present a case of a rare complication of
transvenous RV pacing causing iatrogenic
interventricular septal perforation and
LV pacing wherein a track was formed
leading to frequent displacement of
pacing wire into LV.
Case Presentation
A 63 year old woman
was referred for symptomatic complete heart
block with heart rate of 30 bpm. A
transvenous temporary pacing wire was
inserted into right ventricle (RV) via the
right internal jugular vein. ECG showed
pacemaker spike with good capture but with
right bundle branch block (BBB) pattern and
right axis deviation (Fig.1). Chest X ray (CXR)
showed the wire into left ventricle (LV)
upper border and was posteriorly oriented on
lateral film. A septal perforation with LV
pacing was suspected and the wire was
removed. A new temporary wire was inserted
in our CCU under fluoroscopy through right
femoral vein approach and placed into RV
apex with ECG showing spike with left BBB
pattern.
The patient was asymptomatic but her Troponin T was
elevated. After 2 days, she underwent
coronary angiogram followed by single
chamber permanent pacemaker implantation via
the left subclavian approach with good
position of pacing lead in RV apex confirmed
on multiple fluoroscopy views and on lateral
CXR.
However, the following day after RV pacing insertion,
the ECG again showed right BBB pattern
pacing and the CXR and fluoroscopy showed
high suspicion of lead in LV (Fig. 2&3).
Transthoracic echocardiogram was
inconclusive regarding the lead position and
there was no pericardial effusion. A repeat
ECG with lead placement lower than standard
position showed persistent RBBB. A CT scan
was performed which showed the pacing lead
coursing from RV apex through the
interventricular septum into LV apex
(Fig.4&5). She was taken to the
catheterization laboratory again,
re-exploration was done and the pacing lead
was re-positioned back into the RV free wall
away from apex with good pacemaker position
and parameters. The rest of her stay in the
hospital was uneventful.
Discussion
Temporary transvenous cardiac pacing wires
are usually inserted in an emergency
situation. A recent review of temporary
pacing incorporating 15 studies involving
over 3700 patients from 1973 to 2004 showed
a complication rate of 26.5% and the
commonest complications were failure to
secure venous access, failure to place the
lead correctly, sepsis, puncture of
arteries, lungs or myocardium and
life-threatening arrhythmias. In this
analysis, the incidence of cardiac
perforation was 0.4%1. Acute complications
after permanent pacemaker implantation are
not uncommon, occurring in 4-7% of cases,
and most frequently consist of lead
displacement, traumatic pneumothorax,
hemopneumothorax and pericardial tamponade2.
One potential complication of lead positioning is cardiac
perforation especially with active-fixation
leads, which can cause pericardial effusion,
cardiac tamponade, pneumothorax and death.
Indicators of cardiac perforation include
lead impedance changes, poor sensing or
capture thresholds, diaphragmatic pacing,
right BBB pacing pattern, and/or patient
symptoms of chest pain and hypotension. Most
perforations are extra cardiac with
incidence of 0.4% to 1.2%3,4 and very rarely
occur through the interventricular septum5.
Ventricular pacing usually involves
placement of an electrode in the RV apex and
manifests as left BBB morphology and left
axis deviation on the surface ECG. This
represents a right to left pattern of
depolarization, from the site of electrode
placement i.e., RV apex toward the LV, and
suggests a delay in the depolarization of
the left side on the surface ECG. Right BBB
morphology implies a left-to-right
depolarization pattern and suggests an
incorrect location of the electrode that
initiates the depolarization wave. The
differential diagnosis of paced right BBB
morphology includes inadvertent coronary
sinus placement, inadvertent LV lead
placement, or migration of the electrode
into the LV (through a patent foramen ovale,
atrial septal defect, or septal
perforation). However, normally placed right
ventricular lead in distal septum or apex
(without entering LV) may sometimes produce
right BBB type paced QRS complexes6.
 |
Fig.1: Post-Temporary Pacing ECG
showing pacemaker spike with RBBB
pattern.
|
Pacing electrodes inadvertently placed in the coronary sinus
system can be left alone in asymptomatic
patients, because the depolarization pattern
may resemble physiologic patterns. However,
the presence of pacing electrodes in the LV
is associated with a 37% cerebral embolic
rate, and full-dose anticoagulation is
recommended7. Surgery is also considered for
patients in whom the lead reached the LV as
a result of ventricular septal perforation.
Whether a right BBB pattern induced by ventricular pacing is
the result of a malpositioned lead or an
appropriately positioned right ventricular
pacing lead has to be determined. The usual
modalities to diagnose lead malposition are
traditionally CXR/fluoroscopy along with
echocardiography5 and CT scan8.
In our patient, the echocardiogram was
inconclusive. CXR and fluoroscopy showed a
posteriorly oriented pacing lead which may
also be due to pacing lead in the coronary
sinus. ECG performed one space below showed
persistence of right BBB pattern in V1-2.
Hence, CT scan was done to confirm that the
pacing lead had indeed perforated the septum
and was in the LV apex. The patient was then
taken for re-positioning which was
successfully done without need for surgical
exploration.
If imaging rules out lead malposition, then right BBB pattern
during ventricular pacing may be due to what
is called safe right BBB configuration.
Coman et al.6 reported seven cases with
right BBB pattern during permanent right
ventricular pacing. The placement of leads
V1-V2 one interspace lower than standard
resulted in disappearance of right BBB
morphology and inscription of QS or rS
complexes in V1-V2. The exact mechanism for
this is not known but theories are: portion
of the interventricular septum which are
anatomically RV may behave functionally and
electrically as LV; the right BBB pattern
could be the result of a combination of
right ventricular activation delay due to
severe disease of the right ventricular
conduction system and early penetration of
the electrical impulse into the left
ventricular conduction system.
 |
|
Fig.2: Post-Permanent Pacing lead in
LV on AP view. |
Fig.3: Post -Permanent Pacing lead
in LV on Lateral view. |
To reduce this complication during temporary pacing, there
are some precautions that need to be taken:
1) Caution should be exercised when
utilizing a stiff temporary pacing wire and
minimize the force applied when positioning
the distal tip; 2) avoid excessive turns of
the lead tip to minimize tissue damage and
septal or wall penetration; and 3) to use a
balloon-tipped temporary wires if
available.¨
|
Fig.4: CT Chest showing pacemaker
lead entering from RV into LV. |
|
Fig.4: CT Chest showing pacemaker
lead along the interventricular
septum entering LV.. |
References:
1. McCann P. A Review of Temporary Cardiac
Pacing Wires. Indian Pacing Electrophysiol
J. 2007; 7(1): 40-49.
2. Tobin K, Stewart J, Westveer D, Frumin H.
Acute complications of permanent pacemaker
implantation: their financial implication
and relation to volume and operator
experience. Am J Cardiol 2000; 85:774-776.
3. Sivakumaran S, Irwin ME, Gulamhusein SS,
Senaratne MP. Post pacemaker implant
pericarditis: incidence and outcomes with
active-fixation leads. Pacing Clin
Electrophysiol 2002; 25:833-837.
4. Mahapatra S, Bybee KA, Espinosa RE, Sinak
LJ, McGoon MD, Hayes, DL. Incidence and
predictors of cardiac perforation after
permanent pacemaker placement. Heart Rhythm
2005; 2:907-911.
5. James M, Townsend M, Aldington S. An
unusual complication of transvenous
temporary pacing. Heart 2003; 89:448.
6. Coman JA, Trohman RG. Incidence and
electrocardiographic localization of safe
right bundle branch block configurations
during permanent ventricular pacing. Am J
Cardiol 1995; 76: 781-784.
7. Van Gelder BM, Bracke FA, Oto A, et al.
Diagnosis and management of inadvertently
placed pacing and ICD leads in the left
ventricle: a multicenter experience and
review of the literature. Pacing Clin
Electrophysiol. 2000; 23:877-883.
8. Sunil S, Joshua MC, Albert TC, Michael
AA. Rib Perforation From a Right Ventricular
Pacemaker Lead. Circulation 2007;
115:e391-e392.
THE WHEEL: 3500 BC
Mechanical Perfection
Assyrian King
Assurbanipal on his chariot,
(from 650-659 BC.)
The wheel has been around so long
and is so ubiquitous that it can be
difficult to see exactly why it is
such a brilliant invention. In the
end, it boils down to the way a
wheel exploits the mathematical
properties of a circle.
The definition of a circle is a
curve which keeps the same distance
from a fixed point. So a wheel with
an axle through its center allows
you to keep the same height over the
ground as you roll. And, as tangents
touch circles at only one point, the
rim of a wheel is guaranteed to make
minimal contact with the thing most
likely to slow it down: the ground
[gravity].
Constant height, minimal contact –
geometrical perfection turned into
mechanical perfection.
James Dyson, History of Great
Inventions
|
