TECHNOLOGY
FOCUS: Non-Invasive Coronary Imaging - MRI
MAGNETIC RESONANCE ANGIOGRAPHY IMAGING OF THE CORONARY ARTERIES
Yi Wang Ph.D., H. Dirk Sostman, M.D.
Department of Radiology,, Weill Medical College of Cornell University,
New York, N.Y. , USA
1. CORONARY MOTION
2. EXISTING CORONARY MRA TECHNIQUES
3. CURRENT DEVELOPMENTS IN CORONARY MRA TECHNIQUES
Noninvasive magnetic resonance imaging (MRI) technology has the capacity
to assess cardiac anatomy and function, promising
significant advances in diagnosis and management
of heart disease.
However, as yet the technical and clinical results
from MR imaging specifically of the coronary arteries
have not been sufficient to substantially impact
clinical care of patients with ischemic heart
disease.
Accordingly, an understanding of the imaging approaches,
physical constraints and technical innovations,
which contends with the physiologic and pathologic
substrate of coronary atherosclerosis, is important.
This paper reviews the technical development of
MRI of coronary arteries:
coronary magnetic resonance angiography (MRA).
When suitably developed, coronary MRA will be
the major clinical application of cardiac MRI.
We will summarize the challenges that confront
coronary MRA and the recent developments that
make coronary MRA promising.
We hope to provide the reader an informed appreciation
of how coronary MRA is performed and thus how
it may be used, now and in the foreseeable future.
(Heart Views.2001;2(3):98-111) © 2001 Hamad Medical
Corporation.
Key Words:
magnetic resonance imaging
coronary arteries
atherosclerosis
coronary artery disease
angiography
Cardiovascular disease remains the leading cause
of death in many developed countries.
In the United States, every year, more than 1.
5 million acute myocardial infarctions resulting
in almost 600,000 deaths are attributed to coronary
artery disease (CAD) (1).
Half of these infarctions occur suddenly in previously
asymptomatic individuals (2).
The standard technology for evaluating CAD definitively
is x-ray angiography (XRA), but it is invasive,
expensive, and carries risk.
Cardiac catheterization is associated with 0.12-0.2%
mortality, 0.03-0.2% cerebrovascular accident,
0-0.25% myocardial infarction, 0.57-1.6% local
infarction, 0.3-0.63% arrythmias (3-6).
The cost for coronary XRA is high ($3,500-6,000
in the United States) (7-10).
These limitations make XRA inappropriate for use
as a screening or discovery test.
Accordingly, physicians utilize multiple diagnostic
procedures (including history, blood testing,
stress ECG, coronary calcium scoring, stress ultrasound
and stress perfusion scintigraphy) to assess the
likelihood of CAD in patients.
Despite this typically extensive (and expensive)
clinical, laboratory and imaging evaluation prior
to catheterization, up to 20% of x-ray coronary
angiograms demonstrate no significant coronary
artery disease and therefore do not lead to further
interventions (6).
Furthermore, these pre-angiography tests themselves
carry small risks and expose patients to significant
inconvenience as well as expense.
It would be very desirable to diagnose these stenoses
in a noninvasive and cost effective manner using
MRI.
Patients without evidence of stenosis on noninvasive
imaging would avoid the cost and potential risks
of x-ray angiography
A noninvasive and less expensive diagnostic modality
would advance patient care and might reduce net
healthcare costs for CAD diagnosis if it were
sufficiently accurate and appropriately utilized.
Follow-up imaging of patients undergoing revascularization
procedures could also be readily obtained and
used to guide further therapy. Serial studies
for progression of disease could be carried out,
facilitating the development of drug or gene therapy
for CAD (11).
MRI is a very promising approach for realizing
the potential of noninvasive coronary imaging
because of its high sensitivity to blood flow.
Not only can the vessel lumen be visualized but
also the flow velocity can be mapped directly
and in principle plaque also can be visualized
directly.
MR angiography (MRA) has the potential to image
the proximal coronary arteries (3-5 mm diameter)
and major branches and distal arteries (~2 mm).
The 3D imaging of the coronary tree in MRA would
overcome the vessel-overlapping problem in projectional
XRA.
Clinically significant coronary stenoses (>50%
diameter narrowing) are mostly located in the
proximal portion of the major coronary arteries
(left main, left anterior descending, left circumflex,
right) (12-14). Arteries of this size and smaller
in the brain and extremity are well imaged with
MRA methods, and neuro and peripheral MRA have
become valuable diagnostic tools that are routinely
used in clinical practice (15-18). The major challenge
for MRA of the coronary arteries is cardiac and
respiratory motion.
Initial results using ECG triggering and respiratory
gating or breath-holding demonstrate that coronary
MRA is highly feasible and very promising.
However, as will be described below, further development
is needed to optimize motion suppression and make
the techniques robust for routine clinical application.
In addition to imaging coronary arteries, MRI
is capable of comprehensive study of the anatomy
and function of the heart, so-called “one-stop
shopping” which might include assessing ventricular
function, myocardial perfusion, and wall motion
or thickening.
However, coronary MRA is one of the most important
components of cardiac MRI, and recently has been
identified as the No.
1 special area of emphasis for NIH research (19).
We organize this review article on coronary MRA
development in the following order.
  1) Coronary motion
The motion of coronary arteries, due to respiration and cardiac contraction, is the major challenge to coronary MRA.
Study of the characteristics of coronary motion dissects this major challenge and lays the foundation for coronary MRA techniques.
2) Existing coronary MRA techniques
Techniques that rely on the ECG signal to identify cardiac motion and on diaphragm navigator feedback or breath-holding to minimize respiration effects can depict coronary arteries.
However, image quality is variable and diagnostic accuracy is limited.
3) Coronary MRA techniques under development
A new navigator approach is presented that can directly measure coronary motion and overcome the inaccuracy in detecting coronary motion from the ECG signal and the diaphragm navigator signal.
As mentioned above, the success of a coronary
MRA technique is in large measure determined by
its effectiveness in suppressing motion artifacts,
so the value of studying the characteristics of
coronary motion cannot be overemphasized.
First, we describe the coronary motion due to
respiration.
Then, we examine the coronary motion due to cardiac
contraction.
Of course, cardiac contraction and respiration
occur concurrently and cannot be separated in
motion measurement.
A useful and quantitative mathematical decomposition
of coronary motion is presented.
The coronary motion consists of a dominant global
component that can be measured in MRI and a small
local deformation component.
The heart is situated superior to the diaphragm, which during the respiratory
cycle acts like a piston, pushing and pulling
the heart in the superior-inferior (SI) direction.
Respiration amplitude varies on a 100 msec time
scale and does not have a strictly periodical
waveform, though there is a 2-4 sec, relatively
regular cycle of inspiration and expiration.
The dominant component of coronary motion due
to respiration is a linear translation along the
SI direction, on the order of 10 mm (20).
Other components are much smaller, including translation
along anterior-posterior (AP) and right-left (RL)
directions on the order of 2 mm.
Finally, there is a dilation along the SI direction
because the coronary roots are displaced less
than the distal arteries at the bottom of the
heart.
The coronary motion due to respiration is approximately
correlated to the SI motion of the

Fig.1
 |
Fig.1. An RCA section of the heart at three different respiration level: inspiration (left), middle and expiration (right).
(Courtesy: MRM33:713-719, 1995 )
|
diaphragm, though there may be hysteresis during
the respiratory cycle (21).
These characteristics imply that the respiratory
component of coronary motion can be approximately
monitored at the diaphragm.
However, for maximum accuracy motion measurement
for coronary MRA should be performed directly
on the heart.
The coronary motion due to cardiac contraction and rotation (referred
to as "cardiac motion" in this paper) varies on
a 10 msec (or less) time scale with a fairly periodic
waveform (~ 1 sec cycle).
Although the ECG signal is used to trigger CMRA
acquisition, it is not a measure of coronary motion.
Cardiac motion can be measured quantitatively
using invasive cine x-ray angiography method,
and
Fig.2
 |
Fig.2. Rest period versus heart rate in 13 patients.
As heart rate increases, the duration of
the rest period decreases. The rest periods
for the right and left arteries differed
substantially at low heart rate.
|
properties crucial to CMRA data acquisition have been identified (22).
The rest period or period of minimal cardiac motion
(position variation < 1mm) varies substantially
from patient to patient, and is not uniquely determined
by the length of the cardiac cycle.
The rest period measured from a group of patients
varied from 66 to 333 msec for the left coronary
artery and from 66 to 200 msec for the right coronary
artery.
The cardiac motion velocities measured immediately
after the rest period are significantly larger
than the velocities measured immediately before
the rest period.
The transverse (RL and AP) motion range of the
right coronary artery is greater than that of
the left coronary artery.
When there is no respiratory motion in the thoracic
trunk and no change in cardiac output, the heart
should return to the same location from heartbeat
to heartbeat.
Overall cardiac motion can be decomposed into
two components:
1] global motion (translation, rotation about
the long axis and dilation/contraction at the
center of the heart) and 2] local deformation
that is not accounted for in the global model,
including bulging during contraction and twisting
caused by non-uniform rotation.
A very useful and interesting finding is that
most (~91%) of the 3D cardiac motion of coronary
bifurcations can be represented by the global
motion (23).
The global motion can be measured from its effects
on MR signal.
The local deformation cannot be measured easily,
but its magnitude is proportional to the magnitude
of the global motion.
Accordingly, overall cardiac motion effects can
be reduced by minimizing global motion within
image data. This is the foundation for the use
of navigator measurement of and correction for
global translation, rotation and dilation in MRI.
As mentioned above, cardiac contraction and respiration occur concurrently and cannot be separated in motion measurement.
A useful and quantitative mathematical decomposition of coronary motion is available, however.
The discussion in the above three paragraphs leads to the following mathematical expression for the total global motion of coronary arteries.
Let x be the position vector, and Dr the displacement vector.
Then the respiratory motion component can be expressed in a matrix form:
#x'=Cr(x - Dr),[1]
where Cr is a diagonal matrix characterizing compression/dilation.
Cardiac motion can be expressed conveniently in a coordinate system XYZ, with Z the long axis of the heart and XZ in the plane of the interventricular septum.
The transformation from the laboratory xyz coordinate system to the cardiac major axes XYZ coordinate system consists of 1) rotating about the x-axis until the z-axis becomes the Z-axis and 2) rotating about the Z-axis until the x-axis gets to the septum (XZ) plane (both rotating angles can be easily measured from scout scans of the heart).
This transformation is, of course, intuitively familiar to any trained echocardiographer.
The Jacobian associated with this transformation (J) is the product of these two rotations and can be obtained from a scout scan.
The rotation and contraction associated with cardiac motion, adding the respiratory component, is composed of transforming from xyz to XYZ, displacement (Dc) contraction / dilation (Cc), rotation (Rc), and transforming from XYZ back to xyz:
x"=J-1(RcCcJx'-Dc)=A(x-D)[2]
Coronary arteries can be well visualized using MRI.
The cohort of existing coronary MRA techniques relies on the ECG signal to identify the period of minimal cardiac motion for data acquisition of coronary MRA.
Typically, a delay to mid-diastole estimated from the ECG is used to start MR data acquisition. However, the ECG signal bears neither direct nor quantitative correlation with coronary motion, and consequently there may be substantial residual cardiac motion to degrade image quality.
For suppression of respiration artifacts, breath-holding or navigator technique (see below for a discussion of navigator technique) that measures and corrects for diaphragmatic motion have both been used extensively.
Most coronary MRA techniques use gradient echo signal acquisition in order to acquire ‘bright blood’ image contrast. This has appeal due to similar image contrast with conventional angiography, and in addition it lends itself to further MR signal enhancement from the use of intravenous gadolinium-based MR contrast agents.
Satisfactory image contrast for coronary arteries when using “bright blood” technique depends upon suppression of the bright signal from epicardial fat surrounding the coronary arteries (24).
Image contrast for more distal coronary artery segments can be further enhanced by suppressing the MR signal of surrounding myocardium, which can be achieved using magnetization transfer pulses (25) or T2 preparation pulses (26).
Conventional rectilinear data sampling trajectories in k-space are standard and robust.
More efficient spiral trajectories (27) or multiple rectilinear lines (echo planar imaging) (28) have also been used for data sampling, but substantial diligence is required to minimize off-resonance artifacts.
In summary, a standard data acquisition sequence for coronary MRA consists of 1) magnetization preparation prior to data acquisition commencing at mid-diastole to suppress epicardial fat, and 2) using gradient echo sequence to acquire a few lines in k-space (over a duration of ~ 100 msec).
Depending on the method used to minimize respiratory artifacts, existing coronary MRA techniques can be grouped into breath-holding and navigator types.
The 2D acquisition can be completed within a breath-hold of 20 heartbeats using a gradient echo sequence of TR ~ 10 msec and acquiring 8 k-space lines per heartbeat.
This 2D breath-hold technique was first to demonstrate the feasibility of coronary MRA (24).
However, clinical trials up to date have yielded undesirably variable accuracy (Table 1).
Limitations intrinsic to the breath-hold 2D approach include inconsistency between breath-holds, poor resolution or missing artery segments
in the ‘section’ plane (due to large section thickness and non-isotropic spatial resolution, and poor SNR.
These limitations of 2D data acquisition could be overcome by using a 3D acquisition approach, which provides higher SNR and higher spatial resolution than 2D acquisition and natural 3D display for the tortuous coronary artery tree.
A typical 3D coronary MRA acquisition requires many breath-holds to complete.
It is essential to maintain the heart in the same position during all breath-holds for a multiple breath-hold acquisition.
A respiratory feedback monitor can aid the subject to hold his/her breath consistently (29,30).
A measure of respiration level such as the diaphragm position can be visually displayed to the subject.
While the respiratory feedback can reduce the variation among breath-holds and permit high quality multiple breath-hold 3D coronary MRA (31) (Fig.3), substantial cooperation and comprehension, in addition to simple breath holding, is required from the patient.
This makes such technique a difficult and not very robust choice for coronary imaging in clinical practice. To date, there is no reported clinical result from this technique.
The 3D acquisition can be made fast enough to
allow single breath-hold 3D coronary MRA (32).
This is accomplished by using a high sampling
rate and fast gradient system, but it carries
with it the trade-off of poor SNR.
The relatively poor SNR intrinsic to this approach
can be alleviated by using MR contrast agents
that enhance blood signal substantially.
This approach is used in the vast majority of
clinical MRA in all parts of the body (33).
One specific implementation of such a contrast
enhanced breath-hold 3D coronary MRA technique
is illustrated in Fig.4.
Disdaqs (executing RF and gradient pulses with
data acquisition disabled) and fat saturation
were used to establish spin equilibrium.
All background tissues are well saturated, generating
bright contrast for T1 shortened blood (Fig.5).
During the acquisition window in the cardiac cycle,
an edge-center-edge view order was used such that
the center of k-space was acquired in mid-diastole
when motion of the coronary arteries was minimal.
As is well known, the central k-space views govern
the overall image contrast and provide most of
the signal, so making these correspond to the
motionless portion of the cardiac
Fig.3
 |
Fig. 3. Proximal right coronary artery and left coronary artery demonstrated in an oblique section reformmatted from a 3D data set acquired with multiple breath-holds aided with a respiratory feedback monitor.
|
cycle has obvious advantages (Fig.6).
The whole 3D data acquisition in this specific example was completed in a 32-heartbeat breath-hold.
Data acquisition began when the contrast bolus arrived at the ascending aorta.
Coronary arteries are well delineated with the breath-hold contrast enhanced 3D coronary MRA using motion-matched view ordering, and the background suppression in the contrast enhanced acquisition allows easy display of the 3D coronary data set.
Very likely, contrast enhancement will play a significant role in improving coronary artery contrast.
Initial clinical evaluations of the breath-hold 3D contrast enhanced MRA techniques yielded encouraging results, but the diagnostic accuracy is not yet satisfactory (Table 1).
The data acquisition matrix has to be increased to provide sufficient image resolution, and residual cardiac and respiratory motion effects (diaphragm can drift during breath-holding) have to be reduced.
Breath-hold acquisition has fundamental shortcomings including the necessity
for cooperation in maneuvers that may be beyond
the physiologic capability of certain patients;
limited spatial resolution or SNR due to short
available scan time; and unanticipated or residual
motion during attempts of breath holding.
These drawbacks can be overcome with the navigator
approach.
Fig.4
 |
Fig.4. Timing schematics for contrast enhanced breath-hold 3D coronary MRA using motion matched view order. (Courtesy: Radiology 215:600-607, 2000)
Fig. 5
|
|

|
 |
Fig. 5. (a) an oblique section reformatted
from a breath-hold contrast enhanced 3D
acquisition to demonstrate LAD. (b) is non-enhanced
breath-hold 2D acquisition of the LAD in
the same subject. Compared to (b), the contrast
enhanced acquisition in (a) provided much
better background suppression and correspondingly
substantially improved image coronary artery
contrast. (Courtesy: Radiology 215:600-607,
2000)
|
|
Fig. 6

|

|
Fig.6. Comparing centric (a) to motion matched (b) view order in contrast enhanced 3D CMRA, motion-matched view order substantially sharpens the coronary artery (arrows). (Courtesy: Radiology 215:600-607, 2000)
Fig. 7
|
 |
Fig.7. Timing schematics for navigator gated coronary MRA acquisition.
|
|
Fig.8

|

|
Fig.8. 3D coronary MRA acquired under breathing (a) without navigator gating and (b) with navigator gating. Motion effects (ghosting and blurring) are substantially reduced by navigator gating. Images are reformatted to show the proximal left anterior descending artery (LAD, arrow). (Courtesy: Radiology 198:55-60, 1996)
The concept of navigator motion reduction was introduced in the
early 1950’s for correcting the distorting
effects of the Earth’s atmosphere in ground-based
astronomical imaging (34).
In ground-based optical astronomy, resolution
is limited not only by the size of telescopes,
but importantly by atmospheric turbulence,
which deforms the image on a millisecond
time scale.
Adaptive optical systems consisting of natural
or artificial guide stars, wavefront sensors
and real-time phase delay corrections compensate
for the atmospheric distortions and significantly
improve image resolution (35).
This navigator motion reduction idea has
also been very successfully applied in camcorders
and other imaging technologies (36).
The navigator approach offers a tool to
deal with motion problems in data acquisition
generically.
The key concept is that effects of motion
can be suppressed to a degree determined
by the accuracy achieved in measurement
of the motion.
Spatial resolution of MR images of the moving coronary artery
tree is limited by physiological motion,
not by the intrinsic spatial resolution
of clinical MR imaging (which is determined
by the capability of the MR scanner in sampling
high spatial frequencies).
Navigator based motion gating and motion
correction techniques were introduced in
MRI in the late 1980’s (37,38).
The beauty of navigator technique in MRI
is that the motion to be corrected can be
detected from the MR signal itself, with
no additional hardware needed for motion
monitoring.
Currently navigator motion
Fig.9
|
Fig.9. X-ray angiogram of proximal LAD (a) prior
to angioplasty (b) immediately after angioplasty.
(c) Navigator gated CMRA 5 days after angioplasty,
concordant with x-ray angiogram.
|
correction techniques are widely used to correct undesired motion effects in diffusion imaging, fMRI, body MRA and spectroscopy.
The initial results from our group and others vindicate the utility of these navigator methods for reducing motion effects in CMRA.
The widely used navigator to monitor respiration is the 2D RF excitation pulse (39-41) used to acquire a longitudinal cylinder of tissue through the posterior part of the diaphragm (29).
The diaphragm position can be extracted accurately from the pencil beam navigator echo using edge detection algorithms such as the least squares algorithm (42).
The SI respiratory motion of the heart is then estimated from the SI motion of the diaphragm.
Such a pencil-beam navigator echo can be flexibly positioned away from the heart and thus will not affect the cardiac magnetization.
In addition to monitoring the SI position of the diaphragm, the pencil beam navigator echo can also be used to monitor the AP oriented motion of the chest wall and the upper abdominal wall.
Once the motion of interest is detected, navigator methods to suppress the motion’s undesirable effects on the MR image are applied.
Those explored for CMRA include gating and correction.
The gating method acquires image data only when the amplitude of the detected motion falls within the gating window.
Gating both retrospectively and prospectively has been implemented.
Linear motion can compensated for by phase shifts.
Retrospective navigator gating consists of oversampling data (~5X) and then selecting the set of data with associated diaphragm positions within or closest to the gating window (43).
Because certain data points may lay outside the gating window, this retrospective gating method may be not effective, but it is easy to implement.
Several clinical evaluations of the retrospective gated 3D coronary MRA technique yielded limited accuracy (Table 1).
The effectiveness of navigator gating can be improved using the real time gating, which is the topic of the remaining part of this section and the next section.
Real time gating requires computer capability that can process navigator data and communicate with data acquisition within one R-R interval or less (44,45). . In one implementation of real-time navigator gating for 3D coronary MRA, the pencil beam navigator echoes are acquired immediately before and after the image data in every cardiac cycle (Fig.5).
The SI positions of the diaphragm are extracted from the two navigator echoes in real-time (< 2 ms).
The decision to control data acquisition is made at the beginning of the data acquisition in the next heartbeat.
When both positions are within the gating window, image data collected between the navigator echoes is accepted and image data acquisition is advanced to the next segment of k-space.
Otherwise, the same segment of k-space is reacquired in the next heartbeat.
Such an implementation rejects data if motion occurs during the image data acquisition period.
The location of the gating window can be optimally selected using the diminishing algorithm (46).
The effects of residual motion within the gating window can be minimized by smoothly distributing residual motion in the k-space through view ordering (47).
Real-time navigator gating (at a gating window of 3 mm SI range of the diaphragm) provides significant reduction of respiration effects (Fig.8), and can be used to
follow up patients after revascularization (Fig.9).
Clinical evaluations of the preliminary implementations of the real time navigator gated 3D coronary MRA technique demonstrated variable accuracy (Table 1); the inconsistency in image quality be may caused by the fundamental limitation of the diaphragm navigator echo as discussed in the next session on current developments in coronary MRA techniques.
Effective real-time navigator gating requires a narrow gating window, which leads to undesirable increase in scan time.
To shorten scan time, a wide gating window can be used to increase the data acceptance rate and to correct for the residual displacements within the wide gating window (48-50).
The displacement effects on MRI data are phase shifts, which can be corrected for by multiplying a phase factor:
Fc(k) = F(k) exp[-ik D].[3]
The SI displacement of the diaphragm dSI has to be scaled to estimate the displacement of the heart DSI = a dSI.
The scaling factor a can be searched in the range (0,1] for the best correction (20).
This combination of real-time navigation gating and adaptive motion correction can reduce motion effects without the need for extensive patient cooperation and without an undesirable increase in scan time (Fig. 10).
However, the effectiveness of motion correction based on phase shift is quite limited, because the coronary motion is more complex than the linear motion estimated from the diaphragm.
The inability to achieve consistent image quality through use of existing coronary MRA techniques (as reviewed above) may be due to errors associated with the attempts to estimate the cardiac and respiratory motion components.
The ECG signal is used to minimize cardiac motion, but the ECG has only an indirect and inconsistent correlation with coronary motion.
Breath holding imposes significant limitations on image resolution and/or SNR, and there may still be residual motion despite conscientious effort by the patient .
Diaphragm navigator echo may not accurately monitor cardiac motion due to respiration when there is hysteresis between the diaphragm and heart positions.
Accordingly, most efforts to improve the accuracy of coronary MRA are focusing on achieving more accurate measurement of coronary motion or suppression of motion effects on the MR image.
Pencil beam navigator echo can be positioned directly on the heart, rather than on the diaphragm.
In this way, it is possible to monitor the motion of the heart in a manner similar to an M-mode ultrasound.
Such ‘M-mode’ pencil beam navigator can be used to identify the period of minimal cardiac motion (51).
The motion of coronary arteries is very difficult to extract from such a signal, because it is a mixture of chamber blood, myocardium and arterial blood.
However, if only the coronary artery or the myocardial wall is excited and contributes to the MR signal, then there is no need to use spatial resolution to identify
the heart, and the motion of coronary arteries can be estimated directly and conveniently from the k-space data (52).
An exciting development towards this goal of fast direct and accurate coronary motion measurement is a recently presented cardiac navigator echo using epicardial fat (53).
The heart is comprised of chamber blood, myocardium, and epicardial fat. The motion of coronary arteries correlates directly with the motion of the myocardium and epicardial fat, but not the motion of the chamber blood. Myocardium and blood have the same resonance frequency and therefore it is difficult to excite myocardium only. However, the epicardial fat can be selectively excited using a spatial-spectral selective pulse (53,54). Human coronary arteries are surrounded with plentiful epicardial fat. There is sufficient fat signal for monitoring the coronary motion without affecting coronary blood signal. Preliminary data demonstrate the feasibility of the volumetric fat navigator echo for monitoring cardiac motion (53). This fat navigator echo work suggests the following general volumetric orbital navigator approach to motion issues in coronary MRA.
Note that the global motion of coronary arteries can be characterized by a few parameters (~ 10, Eq.2), and that many points (~ 200) are sampled in an echo in MRI.
So it is possible to solve all global motion parameters from a single navigator echo by designing k-space sampling trajectories that sensitize all motion components.
This leads to the development of orbital navigator echo.
The general k-space expression of the motion effects fm(x) = f(A(x_D)) is:
Fm(k) = eik DF((A-1)Tk) [4]
The six motion parameters in matrix A (the rotation angle in Rc, two scaling
factors in Cc, and three scaling factors in Cr)
can be determined by the least squares minimization,
min{Ýk(|Fm(ATk)|-|F(k)|)2}.
This requires sampling a k-space volume where
ATk might be located around k, and a search for
the minimal by resampling data. Under certain
situations, trajectories may be simplified.
In the 2D case, the k-space trajectory is a circle,
which allows complete determination of rotation
and translation in a plane (52). A trajectory
consisting of 3 axes can be used to simultaneously
detect 3D translation and dilation (53).
Once the global motion is known, its complex effects
on MR signal in Eq.4 can be simultaneously corrected
by adjusting the gradients and RF.
The gradient adjustment according to the following
k-space vector:
Coronary magnetic resonance angiography has advanced substantially in the ten years since it was first attempted.
The progress in coronary MRA has been realized through increased imaging speed due to more powerful gradient hardware, implementation of blood signal enhancement through use of Gadolinium contrast agents, better understanding of coronary motion resulting in development of techniques for more effective suppression of motion effects in coronary MRA data acquisition.
Coronary MRA has been found useful clinically in checking the patency of coronary by-pass grafts (55,56) and identifying anomalous coronary arteries (57,58).
In spite of these advances, consistent and accurate imaging coronary lesions of clinical significance remains elusive, as reviewed in this paper. The fundamental limitation common to all existing coronary MRA techniques is that the motion of coronary arteries during data acquisition is not measured accurately enough (or not measured at all). The errors in motion measurement are the bottleneck for effectiveness of motion suppression in coronary MRA. Once coronary motion is accurately measured, the real-time navigator approach can suppress motion effects in coronary MRA, limited primarily by the
accuracy of the motion measurement.
|
|
Technique
|
Ref #
|
n
|
Sensitivity
|
Specificity
|
Comments
|
Breath-hold 2D
|
60
|
39
|
90
|
92
|
overall
|
100
|
100
|
LM
|
87
|
92
|
LAD
|
71
|
90
|
LCX
|
100
|
78
|
RCA
|
61
|
20
|
63
|
56
|
overall
|
50
|
84
|
LM
|
73
|
37
|
LAD
|
0
|
82
|
LCX
|
62
|
56
|
RCA
|
|
|
62
|
39
|
Na
|
Na
|
|
|
63
|
35
|
100
|
93
|
LM
|
53
|
73
|
LAD
|
0
|
96
|
LCX
|
71
|
82
|
RCA
|
|
64
|
108
|
85
|
80
|
Severe
stenosis
|
38
|
83
|
Moderate
stenosis
|
Breath-hold
|
65
|
11
|
Na
|
Na
|
|
contrast
|
66
|
50
|
94.4
|
57.1
|
overall
|
enhanced 3D
|
|
|
100
|
97
|
LM
|
94
|
91
|
Proximal LAD
|
100
|
91
|
mid LAD
|
50
|
84
|
Proximal LCx
|
67
|
94
|
mid LCx
|
92
|
88
|
Proximal RCA
|
75
|
89
|
mid RCA
|
|
67
|
38
|
68
|
97
|
overall
|
77
|
97
|
LM+LAD
|
50
|
100
|
LCX
|
64
|
94
|
RCA
|
Navigator
gated
3D - retrospective
|
68
|
83
|
83
|
94
|
Significant stenosis
|
Na
|
96
|
LM
|
67
|
72
|
Proximal LAD
|
71
|
75
|
LCX
|
66
|
90
|
Proximal RCA
|
|
69
|
35
|
75
|
100
|
LM
|
87
|
85
|
Proximal LAD
|
88
|
90
|
LCX
|
50
|
95
|
Proximal RCA
|
|
|
70
|
10
|
73
|
Na
|
|
|
|
71
|
29
|
78
|
86
|
overall
|
|
|
72
|
42
|
82
|
89
|
overall
|
|
|
|
|
90
|
90
|
Proximal segments
|
68
|
81
|
Distal segments
|
Navigator gated
3D - real time
|
73
|
20 |
65
|
93
|
overall
|
NA
|
93
|
LM
|
100
|
86
|
LAD
|
33
|
100
|
LCX
|
71
|
90
|
Proximal
RCA |
|
|
74
|
7
|
Na
|
Na
|
|
Abbreviations used: Ref # = the reference number
of the published study (2nd column),
n = the number of patients enrolled
in a study (3rd column), LM = left
main artery, LAD = left anterior descending
artery, LCx = left circumflex artery,
RCA = right coronary artery, Na =
not available.
|
|
Current work on directly measuring the motion of coronary arteries may lead to accurate coronary motion measurement (53) and may provide a potential solution to the fundamental motion problem of coronary MRA.
On-going work in fast imaging, such as true FISP imaging (59), may further increase data acquisition speed and signal-to-noise ratio.
These current developments in coronary motion measurement and fast imaging are encouraging.
They certainly will bring us closer to the goal of accurate imaging of coronary lesions, and may allow us to reach this goal in the foreseeable future.
1. American Heart Association. Heart and stroke facts:
statistics. 1993;8. Dallas,
Texas.
2. Fuster V, Badimon L, Cohen M, et
al. Insights in to the pathogenesis of acute
ischemic
syndromes. Circulation 1998;77:1213-1220.
3. Davis K, Kennedy JW, Kemp HG, Judkins
MP, Gosselin AJ, Killip T.
Complications of coronary arteriography from the collaborative
study of
coronary artery surgery
*CASS). Circulation
1979;59:1105-1112.
4. Kennedy JW. Complications
associated with cardiac
catheterization
and anigraphy. Cathet Cardiovasc Diagn 1982;8:5-11.
5. Wyman RM, Safian RD, Portway
V, Skillman JJ, MacKay RG, Baim DS. Current
complications
of diagnostic and therapeutic cardiac catheterization.
J Am Coll Cardiol 1988;12:1400-1406.
6. Johnson LW, Lozner EC, Johnson
S, et al. Coronary arteryography 1984-1987: a
report
of the Registry of the Society for Cardiac Angiography
and
Interventions. I.Results and complicatoins. Cathet
Cardiacvasc
Diagn.1989;17:5-10.
7. Hansing CE. The risk and
cost of coronary angiography.I. Cost of coronary
angiography
in Washington State. JAMA. 1979;242(8):731-4.
8. Thomson PD. Cost of coronary
angiography. JAMA.1980;243(12):1232-3
9. Keavney B. Haider YM. McCance
AJ. Skehan JD. Normal coronary angiograms:
financial
victory from the brink of clinical defeat? Heart.
1996;75(6):623-5.
10. Manning WJ, Edelman RR. Magnetic resonance coronary angiography. Magn.
Reson. Quarterly 1993;9:131-151,.
11. Mack CA. Magovern CJ. Budenbender KT. et al.Salvage angiogenesis
induced by
adenovirus-mediated
gene transfer of vascular endothelial growth factor
protects
against ischemic
vascular occlusion. J Vasc Surg. 1998;27(4):699-709.
12. Vieweg WV. Alpert JS. Johnson
AD. Dennish
GW. Nelson DP. Warren SE. Hagan AD. Distribution
and severity of
coronary
artery disease in 500 patients with angina pectoris.
Catheterization &
Cardiovascular
Diagnosis. 1979;5(4):319-30.
13. Vieweg WV. Alpert JS. Hagan AD. Caliber
and distribution of
normal coronary arterial
anatomy.
Catheterization & Cardiovascular Diagnosis.1976;2(3):269-80.
14. MacAlpin RN. Abbasi AS. Grollman JH
Jr. Eber L.
Human coronary artery size during
life.
A cinearteriographic
study. Radiology 1973;108(3):567-76.
15. Potchen, E.M. Haacke, J.E. Siebert,
A. Gottschalk, “Magnetic resonance
angiography:
concepts & applications”, Mosby-Year Book Inc.,
St. Louis, 1993.
16. Anderson, R.R. Edelman, P.A. Turski,
“Clinical magnetic resonance angiography”,
Raven Press,
New York, 1993.
17. Baum RA. Rutter CM. et al.
Multicenter
trial to evaluate vascular magnetic resonance
angiography
of
the lower extremity. American College of Radiology
Rapid
Technology
Assessment Group. JAMA. 1995;274(11):875-80
18. Owen RS. Carpenter JP. Baum RA. Perloff
LJ. Cope C.
Magnetic resonance
imaging of angiographically occult runoff vessels
in peripheral
arterial occlusive
disease. New England Journal of Medicine. 1992.326(24):1577-81..
19. Budinger TF, Berson A, McVeigh ER, Pettigrew
RI, Pohost GM,
Watson JT, Wickline
SA.
Cardiac MR imaging: report of a Working Group
sponsored by the
National Heart,
Lung, and Blood Institute. Radiology 1998;208:573-576.
20.Wang Y, Riederer SJ, Ehman RL, Respiratory
motion of the heart: kinematics and the implications
for the spatial resolution in coronary imaging,
Magn Reson Med
1995;33:713-719,.
21. Nehrke K, Bornert P. Study of the respiratory
motion of the heart using multiple
navigator
pulses. ISMRM, 404,200.
22. Wang Y, Vidan E, Bergman GW. Cardiac
motion of coronary arteries: variability in
the rest
period and implications for coronary MR
angiography.Radiology.
1999;
213 (3):751-8.
23. Potel MJ, Rubin JM, Mackay SA, et al.
Methods for evaluating cardiac wall motion in
three dimensions
using bifurcation points of the coronary arterial
tree. Invest Radiol
18: (1) 47-57
1983.
24. Edelman RR, Manning WM, et al, Coronary
arteries: breath-hold MR angiography.
Radiology.
1991;181:641-643.
25. Wang Y, Grist TM, Mistretta CA, Dispersion
in magnetization transfer contrast at a
given specific
absorption rate due to variations of RF pulse
parameters in the
magnetization
transfer preparation, Magnetic Resonance
in Medicine .1997;37:957-962.
26. Brittain JH, Hu BS, Wright GA, Meyer
CH, Macovski A, Nishimura DG. Coronary
angiography
with magnetization-prepared T2 contrast. Magn
Reson Med.
1995;33(5):689-96.
27. Meyer CH, Hu BS, Nishimura DG, Macovski
A. Fast spiral coronary artery imaging.
Magn Reson
Med. 1992;28(2):202-13.
28. Wielopolski PA, Manning WJ, Edelman
RR. Single breath-hold volumetric imaging of
the heart
using magnetization-prepared 3-dimensional segmented
echo planar
imaging.
J Magn Reson Imaging.1995;5(4):403-9.
29. Liu Y, Riederer SJ, et al, A monitoring,
feedback, and triggering system for
reproducible
breath-hold MR imaging. Magn Reson Med 1993;30:507-511.
30. Wang Y, Christy PS, Korosec FR, Alley MT,
Grist TM, Polzin JA, Mistretta CA.
Coronary MRI
with a respiratory feedback monitor: the 2D imaging
case.
Magn Reson
Med. 1995;33(1):116-21.
31. Wang Y, Grimm RC, Rossman PJ, Riederer
SJ, Ehman RL, 3D coronary MR
Angiography
in multiple breath-holds using a
respiratory
feedback monitor, Magnetic Resonance in Medicine.
1995;34, 11-16.
32. Wielopolski PA, van Geuns RJ, de Feyter
PJ, Oudkerk M. Breath-hold coronary MR
angiography
with volume-targeted imaging. Radiology. 1998;209(1):209-19.
33. Prince MR, Grist TM, Debatin JF. 3D
contrast MR angiography. Springer, -Verlag,
Berlin, 1999.
34. HW Babcock, The possibility of compensating
astronomical seeing, Publ. Astron.
Soc. Pac.
65, 229,1953.
35. Thompson LA, Adaptive optics in astronomy.
Physics Today. 1994;27(12):24-31.
36. Shake-free video, Popular Science, p48,
September 1998..
37. RS Hinks, Monitored echo gating (mega)
for the reduction of motion artifacts, Proc.
6th SMRI,
p.48, 1988.
38. RL Ehman, JP Felmlee, Adaptive technique
for high-definition MRI of moving
structures.
Radiology 1989;173:255-263,.
39. Pauly J, Nishimura D, Macovski A. A
k-space analysis of small-tip-angle excitation.
J. Magn.
Reson. 1989;81, 43-56.
40. Pauly J, Nishimura D, Macovski A. A
linear class of large- tip-angle selective
excitation
pulses. J. Magn. Reson. 1989;82, 571-587.
41. Hardy CJ, Cline HE. Broadband nuclear
magnetic resonance pulsese with two-
dimensional
spatial selectivity. J. Appl. Phys. 1989;66, 1513-1516.
42. Wang Y, Grimm RC, Felmlee JP, Riederer
SJ, Ehman RL. Algorithms for extracting
motion information
from navigator echoes. Magn Reson Med. 1996;36(1):117-23.
43. Li D. Kaushikkar S. Haacke EM. Woodard
PK. Dhawale PJ.
Kroeker RM.
Laub G. Kuginuki Y. Gutierrez FR.Coronary arteries:
three-dimensional
MR imaging with
retrospective
respiratory gating. Radiology. 1996;201(3):857-63.
44. TS Sachs, GH Meyer, et al, Real-time
motion detection in spiral MRI using
navigators.
Magn Reson Med 1994;32:639-645.
45. Y Wang, PJ Rossman, et al, Navigator-based
real-time respiratory gating and
triggering
for reduction of respiration effects in three-dimensional
coronary MR
imaging.
Radiology 1996;198:55-60,.
46. Sachs TS, Meyer CH, Irarrazabal P, Hu
BS, Nishimura DG, Macovski A. The
diminishing
variance algorithm for real-time reduction of
motion artifacts in MRI.
Magn Reson
Med. 1995;34(3):412-22.
47. Jhooti P, Gatehouse PD, Keegan J, Bunce
NH, Taylor AM, Firmin DNPhase ordering
with automatic
window selection (PAWS): a novel motion-resistant
technique for 3D
coronary
imaging. Magn Reson Med. 2000;43(3):470- 80.
48. Oshinski JN. Hofland L. Mukundan S Jr.
Dixon WT. Parks WJ. Pettigrew RI. Two-
dimensional
coronary MR angiography without breath holding.
Radiology.
1996;201(3):737-43.
49. P.G. Danias, M.V. McConnell, V.C. Khasgiwala,
M.L. Chuang, R.R. Edelman, W.J.
Manning.
Prospective navigator correction of image position
for coronary MR
angiography.
Radiology 1997;203, 733-736.
50. Wang Y, Ehman RL. Retrospective adaptive
motion correction for navigator-gated
3D coronary
MR angiography. J Magn Reson Imaging. 2000;11(2):208-14.
51. Wang Y, Watts R, Mitchell I, Nguyen
TD, Bezanson JW, Bergman GW, Prince MR.
Coronary
MR angiography: selection of acquisition window
of
minimal cardiac motion
with electrocardiography-triggered navigator cardiac
motion prescanning--initial
results.
Radiology. 2001;218(2):580-5.
52. Fu ZW, Wang Y, Grimm RC, Rossman PJ,
Felmlee JP, Riederer SJ, Ehman RL.
Orbital navigator
echoes for motion measurements in magnetic resonance
imaging.Magn
Reson Med. 1995;34(5):746-53.
53. Nguyen TD, Wang Yi, Prince MR, A Novel
Navigator Technique for Fast and Direct
Detection
of 3D Displacement of the Coronary Arteries, ISMRM
173,2001.
54. Meyer CH. Pauly JM. Macovski A. Nishimura
DG. Simultaneous spatial and spectral
selective
excitation. Magnetic Resonance in Medicine. 1990;15(2):287-304.
55. Wintersperger BJ, Engelmann MG, von
Smekal A, Knez A, Penzkofer HV,
Hofling B,
Laub G, Reiser
MF. Patency of coronary bypass grafts:
assessment
with breath-hold contrast-enhanced MR angiography--value
of a non-electrocardiographically
triggered
technique. Radiology.
1998;208(2):345-51.
56. Langerak SE, Kunz P, de Roos A, Vliegen
HW, van Der Wall EE. Evaluation of
coronary
artery bypass grafts by magnetic resonance
imaging.
J Magn Reson Imaging.1999;10(3):434-41.
57. Duerinckx AJ, Bogaert J, Jiang H, Lewis
BS. Anomalous origin of the left coronary
artery: diagnosis
by coronary MR angiography. Am J Roentgenol.
1995; 164(5):1095-7.
58. McConnell MV, Stuber M, Manning WJ.
Clinical role of coronary magnetic resonance
angiography
in the diagnosis of anomalous coronary arteries.
J Cardiovasc Magn
Reson. 2000;2(3):217-24.
59. Deshpande VS, Shea SM, Laub G, Simonetti
OP, Finn JP, Li D. 3D magnetization-
prepared
true-FISP: A new technique for imaging coronary
arteries.
Magn Reson
Med. 2001;46(3):494-502.
60. Manning WJ, Li W, Edelman RR. A preliminary report comparing magnetic
resonance
coronary angiography with conventional angiography.
N Engl J Med.
1993; 328(12):828-32.
61. Duerinckx AJ, Urman MK. Two-dimensional
coronary MR angiography: analysis of
initial clinical
results. Radiology. 1994;193(3):731-8.
62. Pennell DJ, Bogren HG, Keegan J, Firmin
DN,Underwood SR. Assessment of
coronary
artery stenosis by magnetic resonance imaging.
Heart. 1996;75(2):127-33.
63. Post JC, van Rossum AC, Hofman MB, de
Cock CC,Valk J, Visser CA.
Clinical utility of two-dimensional magnetic resonance angiography
in detecting coronary artery disease.
Eur Heart J. 1997;18(3):426-33.
64. Watanuki A, Yoshino H, Udagawa H, Yokoyama
K, Nitatori T, Hachiya J, Ishikawa K.
Quantitative
evaluation of coronary stenosis by coronary magnetic
resonance
angiography.
Heart Vessels. 2000;15(4):159-66.
65. Kessler W, Laub G, Achenbach S, Ropers
D, Moshage W, Daniel WG. Coronary
arteries:
MR angiography with fast contrast-enhanced three-dimensional
breath-hold
imaging--initial
experience. Radiology. 1999;210(2):566-72.
66. Regenfus M, Ropers D, Achenbach S, Kessler
W, Laub G, Daniel WG,
Moshage W. Noninvasive detection of coronary artery stenosis
using contrast-enhanced three- dimensional
breath-hold magnetic resonance
coronary angiography. J Am Coll
Cardiol. 2000;36(1):44-50.
67. van Geuns RJ, Wielopolski PA, de Bruin
HG, Rensing BJ, Hulshoff M, van Ooijen
PM, de Feyter
PJ, Oudkerk M. MR coronary angiography with breath-hold
targeted
volumes:
preliminary clinical results. Radiology.2000;217(1):270-7.
68. Kessler W, Achenbach S, Moshage W, Zink
D, Kroeker R, Nitz W, Laub G,
Bachmann
K. Usefulness of respiratory gated magnetic resonance
coronary
angiography
in assessing narrowings > or = 50% in diameter
in native coronary
arteries
and in aortocoronary bypass conduits. Am J Cardiol.
1997;80(8):989-93.
69. Muller MF, Fleisch M, Kroeker R, Chatterjee
T, Meier B, Vock P.
Proximal coronary
artery stenosis:
three-dimensional MRI with fat saturation and
navigator echo.
J Magn
Reson Imaging. 1997;7(4):644-51.
70. Woodard PK, Li D, Haacke EM, Dhawale PJ,
Kaushikkar S, Barzilai B,
Braverman AC,
Ludbrook PA, Weiss AN, Brown JJ, Mirowitz SA,
Pilgram TK, Gutierrez FR. Detection of
coronary stenoses on source and
projection images using three-dimensional MR
angiography
with retrospective respiratory gating: preliminary experience.
Am J Roentgenol. 1998;170(4):883-8.
71. Schmidt M, Theissen P, Crnac J, Jochims M, Voth E, Baer FM, Schicha H, Erdmann E.
[Clinical value of magnetic resonance tomography in imaging coronary stenoses.
A comparison
with coronary angiography and myocardial scintigraphy].
Dtsch Med Wochenschr. 1999;124(44):1294-300. German.
72. Sardanelli F, Molinari G, Zandrino F, Balbi M. Three-dimensional, navigator-echo MR
coronary angiography in detecting stenoses of the major epicardial vessels, with
conventional
coronary angiography as the standard of reference. Radiology.
2000; 214(3):808-14.
73. Lethimonnier F, Furber A, Morel O, Geslin P, L'Hoste P,Tadei A,
Jallet P, Caron-Poitreau C, Le Jeune JJ. Three-dimensional coronary artery
MR imaging using prospective real-time respiratory navigator and linear phase
shift processing: comparison with conventional coronary
angiography.
Magn Reson Imaging.1999;17(8):1111-20.
74. Stuber M, Botnar RM, Danias PG, Sodickson DK, Kissinger KV, Van Cauteren M, De
Becker J, Manning WJ. Double-oblique free-breathing high resolution three-dimensional
coronary magnetic
resonance angiography. J Am Coll Cardiol. 1999;34(2):524-31.
Interior of The Great Mosque of Cordoba
Cordoba, Spain
|
|
|
|
|