Vol.14 /No: 2/ Nov 2005

 

   

 

 

Pulse Oximetry: A Useful Tool

Ramaiah V.K..
Department of Anesthetics, Morriston Hospital
Morriston, Swansea, UK
 

Introduction
Principles of Pulse Oximetry
Two Important Laws
   

Introduction:

The principal advantage of optical sensors for medical applications is their intrinsic safety since there is no electrical contact between the patient and the equipment.

Pulse oximetry is a simple, continuous, non-invasive method of monitoring the percentage of haemoglobin (Hb) which is saturated with oxygen. Oxygen saturation (SaO2) is defined as the ratio of oxygen content of hemoglobin over oxygen carrying capacity. A pulse oximeter consists of two light-emitting diodes, a photocell detector and a microprocessor with a visual display unit. The unit displays the percentage of Hb saturated with oxygen together with a signal that is audible for each pulse beat, a calculated heart rate and in some models, a graphical representation of the blood flow past the probe. An oximeter gives early warning of hypoxic events.
 

Principles of Pulse Oximetry:

Oximeters work on the principle of light absorption. Spectrophotometric analysis is the principle of shining radiation through a sample and determining the quantity of radiation absorbed.


There are two important laws:

Beer’s law: Absorption of a given thickness of a solution of a given concentration is the same as twice the thickness of half the concentration.

Lambert’s law: Each layer of equal thickness absorbs an equal fraction of radiation which passes through it.

The wavelengths of light chosen are (660 nm and 940 nm) since this gives a better separation of wavelengths. An isobestic point is the point at which two substances absorb a certain wavelength of light to the same extent. The light is partly absorbed by hemoglobin in an amount that differs depending on whether or not the Hb is saturated with oxygen. By calculating the absorption at the two wavelengths the processor can compute the proportion of hemoglobin which is oxygenated. The computer within the oximeter is capable of distinguishing pulsatile (AC component) flow from other more static signals, such as tissue or venous signals (DC component) to display only the arterial flow. Diodes are switched on in sequence with a pause where both diodes are off. This allows the microprocessor to compensate for ambient light. The diodes are switched off and on in sequence hundreds of times a second – thus the processor can detect cyclical changes due to arterial blood flow. The non-pulsatile component is disregarded.

These oximeters are accurate in the range of oxygen saturation 70 to 100% (+/2%), but less accurate under 70%. The pitch of the audible pulse signal falls with reducing values of saturation.

In the following situations the pulse oximeter readings might not be accurate:

1. Peripheral vasoconstriction (hypovolemia, severe hypoten- sion, cold, cardiac failure, some cardiac arrhythmias) or peripheral vascular disease. This is due to an inadequate signal.

2. Venous congestion, especially when tricuspid regurgitation produces venous pulsations.

3. A badly positioned probe.

4. Bright overhead lights in the theatre might cause the oximeter to be inaccurate (Ambient light).

5. The signal might be interrupted by surgical diathermy.

6. Shivering can cause difficulties in picking up the signal.

7. Scattering and refraction of light.

8. Dyes, nail varnish, patient movement, methylene blue.

9. Tendency to over read in the presence of carboxyhemoglobin and under read in the presence of methemoglobin.

Uses of pulse Oximetry:

1. Forms part of the minimum mandatory monitoring in an operating theatre.

2. Management and weaning from ventilators in an Intensive Care Unit.

3. Procedures done under sedation.

4. Casualty.

Pulse oximetry is possibly the greatest advance in patient monitoring for many years and it is hoped that eventually its use during anesthesia and surgery will become routine worldwide.

CONTINUOUS MEDICAL EDUCATION